7 TROUBLESHOOTING/MAINTENANCE - USER/DEALER SYMPTOM PROBABLE CAUSE USER SOLUTIONS DEALER SOLUTIONS • Bed sections do not move. • Pull tube. • Check pull tube(s) for binding. Call dealer for repairs. • Ensure pull tubes are properly connected. Refer to Adjusting and Reconnecting the Head Section Pull Tube on page 47. • Bed ends move in opposite directions. • Drive Shaft connected to wrong gear box shaft. • Call dealer. • Reinstall Drive Shaft. Refer to Assembling the Drive Shaft on page 51. • Hi/Lo function does not work. • Drive Shaft disconnected. • Call dealer. • Reinstall Drive Shaft. Refer to Assembling the Drive Shaft on page 51. • Pressing function on pendant • Motor cable connections are • Call dealer. results in motion other than that incorrect. intended. • Verify Connections. Refer to Inspecting the Bed on page 54.  For access to part numbers, see online parts book at www.invacare.com or call Invacare. Part No 1180159-E 63 Invacare® G5510 Bed

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69

7
TROUBLESHOOTING/MAINTENANCE - USER/DEALER
Part No 1180159-E
63
Invacare
®
G5510 Bed
Bed sections do not move.
Pull tube.
Check pull tube(s) for binding. Call
dealer for repairs.
Ensure pull tubes are properly
connected. Refer to Adjusting
and Reconnecting the Head
Section Pull Tube
on page 47.
Bed ends move in opposite
directions.
Drive Shaft connected to
wrong gear box shaft.
Call dealer.
Reinstall Drive Shaft. Refer to
Assembling the Drive Shaft
on
page 51.
Hi/Lo function does not work.
Drive Shaft disconnected.
Call dealer.
Reinstall Drive Shaft. Refer to
Assembling the Drive Shaft
on
page 51.
Pressing function on pendant
results in motion other than that
intended.
Motor cable connections are
incorrect.
Call dealer.
Verify Connections. Refer to
Inspecting the Bed
on page 54.
For access to part numbers, see online parts book at www.invacare.com or call Invacare.
SYMPTOM
PROBABLE CAUSE
USER SOLUTIONS
DEALER SOLUTIONS