Troubleshooting/maintenance - User/dealer, Symptom, Probable Cause, User Solutions, Dealer Solutions
View all Invacare G5510 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 62 highlights
SYMPTOM • Bed section does not move. • Bed section does not stop moving. • Bed Hi/Lo movement does not stop. 7 TROUBLESHOOTING/MAINTENANCE - USER/DEALER PROBABLE CAUSE USER SOLUTIONS DEALER SOLUTIONS • Broken connector cap. • Call dealer. • Call Invacare for instructions to replace connector cap. • Motors are not functioning. • Call dealer. • Check to make sure nothing is blocking actuator. • Hi/Lo Motor not functioning. • Ensure cables are not entangled or pinched. • Call dealer. • Ensure cables are not entangled or pinched. • Ensure Drive Shaft is correctly installed. • Ensure Hi/Lo Motor cable is correctly connected. Refer to Assembling the Drive Shaft on page 51. • Thermal Protection Activated. • Allow bed unit to sit for up to 30 • Allow bed unit to sit for 30 minutes and retry. minutes and retry. • Pendant. • Check for depressed/stuck buttons. If button is stuck, call dealer for repair. Unplug power cord from the wall. • Replace pendant. Refer to Electrical Connections on page 55. • Defective Pendant. • Damaged Cables. • Check cables, if damaged call dealer. • Check cables, if damaged, replace. Refer to Assembling the Drive Shaft on page 51. Part No 1180159-E 61 Invacare® G5510 Bed