Invacare JASMINE Owners Manual - Page 20
Troubleshooting
View all Invacare JASMINE manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 20 highlights
SECTION 6-TROUBLESHOOTING SECTION 6-TROUBLESHOOTING SYMPTOMS Patient Lift feels loose. Noisy or dry sound from pivots. Electric actuator fails to lift when button is pressed. FAULTS Mast/Base joint loose. Tie - Rods are loose. Needs lubrication. Hand-control or actuator connector loose. Battery low. SOLUTION Refer to Replace the locknut. on page 26. Refer to Lubricating the Lift on page 22. Charge batteries. Refer to Charging the Battery on page 13. Rotate RED emergency stop button RED emergency stop button pressed IN. CLOCKWISE until it pops out. Battery not connected properly to control box. The connecting terminals are damaged. Reconnect the battery to the control box. Refer to Charging the Battery on page 13. Replace the battery pack. Refer to Charging the Battery on page 13. Electric actuator in need of service or load is too high. Refer to Replacing the Boom Actuator on page 22. Contact your Dealer. Unusual noise from actuator. Actuator is worn or damaged or spindle Refer to Replacing the Boom Actuator on is bent. page 22. Lift arms will not lower in uppermost position. Lift arms require a minimum weight load Pull down slightly on the lift arms. to lower from the uppermost position. Lift arms will not lower during a power retraction. Shoulder bolt at the junction of the boom Refer to Replacing the Swivel Bar or and mast may not be properly installed. Optional Cradle Attachment on page 25. NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare. Jasmine™ Patient Lift 20 Part No 1150704