Invacare JASMINE Owners Manual - Page 20

Troubleshooting

Page 20 highlights

SECTION 6-TROUBLESHOOTING SECTION 6-TROUBLESHOOTING SYMPTOMS Patient Lift feels loose. Noisy or dry sound from pivots. Electric actuator fails to lift when button is pressed. FAULTS Mast/Base joint loose. Tie - Rods are loose. Needs lubrication. Hand-control or actuator connector loose. Battery low. SOLUTION Refer to Replace the locknut. on page 26. Refer to Lubricating the Lift on page 22. Charge batteries. Refer to Charging the Battery on page 13. Rotate RED emergency stop button RED emergency stop button pressed IN. CLOCKWISE until it pops out. Battery not connected properly to control box. The connecting terminals are damaged. Reconnect the battery to the control box. Refer to Charging the Battery on page 13. Replace the battery pack. Refer to Charging the Battery on page 13. Electric actuator in need of service or load is too high. Refer to Replacing the Boom Actuator on page 22. Contact your Dealer. Unusual noise from actuator. Actuator is worn or damaged or spindle Refer to Replacing the Boom Actuator on is bent. page 22. Lift arms will not lower in uppermost position. Lift arms require a minimum weight load Pull down slightly on the lift arms. to lower from the uppermost position. Lift arms will not lower during a power retraction. Shoulder bolt at the junction of the boom Refer to Replacing the Swivel Bar or and mast may not be properly installed. Optional Cradle Attachment on page 25. NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare. Jasmine™ Patient Lift 20 Part No 1150704

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68

SECTION 6—TROUBLESHOOTING
Jasmine™ Patient Lift
20
Part No 1150704
SECTION 6—TROUBLESHOOTING
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare
.
SYMPTOMS
FAULTS
SOLUTION
Patient Lift feels loose.
Mast/Base joint loose.
Tie - Rods are loose.
Refer to Replace the locknut.
on page 26.
Noisy or dry sound from pivots.
Needs lubrication.
Refer to Lubricating the Lift
on page 22.
Electric actuator fails to lift when button
is pressed.
Hand-control or actuator connector
loose.
Battery low.
RED emergency stop button pressed IN.
Battery not connected properly to con-
trol box.
The connecting terminals are damaged.
Electric actuator in need of service or
load is too high.
Charge batteries. Refer to Charging the
Battery
on page 13.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the control box.
Refer to Charging the Battery
on page 13.
Replace the battery pack. Refer to Charg-
ing the Battery
on page 13.
Refer to Replacing the Boom Actuator
on
page 22. Contact your Dealer.
Unusual noise from actuator.
Actuator is worn or damaged or spindle
is bent.
Refer to Replacing the Boom Actuator
on
page 22.
Lift arms will not lower in uppermost
position.
Lift arms require a minimum weight load
to lower from the uppermost position.
Pull down slightly on the lift arms.
Lift arms will not lower during a power
retraction.
Shoulder bolt at the junction of the boom
and mast may not be properly installed.
Refer to Replacing the Swivel Bar or
Optional Cradle Attachment
on page 25.