Campbell Scientific AVW200 AVW200-series Vibrating Wire Interfaces - Page 81
Control Parameters When Troubleshooter is Running, Appendix D., The Public Table
View all Campbell Scientific AVW200 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 81 highlights
Appendix D. The Public Table Control Parameters When Troubleshooter is Running TimeSeries(1) Writing this variable will force a Vibrating Wire measurement and create the TimeSeries.bin and spectrum.bin files. Example: 101 = measures AVW200 chan1 and Mux chan1 102 = measures AVW200 chan1 and Mux chan2 ..... 201 = measures AVW200 chan2 and Mux chan1 202 = measures AVW200 chan2 and Mux chan2 .... 232 = measures AVW200 chan2 and Mux chan32 Notes: TimeSeries(2)..(4) need to be filled with valid values before making the measurement. TimeSeries(2) Begin Frequency for force measurement (100 - 6500) TimeSeries(3) Eng Frequency for forcing measurement (100 - 6500) TimeSeries(4) Excite voltage for forcing measurement (1=5volts or 2=12volts) TimeSeries(5) Copy of the Chan/MuxChan executed in TimeSeries(1) TimeSeries(6) Measured Frequency (Hz) TimeSeries(7) Measured Amplitude (MVolts_RMS) TimeSeries(8) Measured Signal-to-Noise Ratio TimeSeries(9) Measured Noise Freq (Hz) TimeSeries(10) Measured Decay Ratio TimeSeries(11) Measured Thermistor (ohms) D-3