AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules FIGURE 5-2. Deployment Communications Editor in DevConfig NOTE Certain AVW206 settings must match the RF401 settings for communications between the interface and radio to be successful. Description of the Communication Settings follows: Protocol-choose "PakBus" for the "Protocol" setting. Please note that the "Protocol" setting for the RF401-series radio must be set to either "PakBus Aware" or "PakBus Node" for communications to be successful. PakBus Address-enter a PakBus Address that matches the PakBus address specified in the CRBasic program that will control the AVW200. Each device in a PakBus network must have a unique address. Valid settings are 1 - 4094. RS-232 Baud Rate-enter the baud rate in which you want to communicate. Hop Sequence-enter the radio "Hop Sequence" that matches all of the RF401 radios and other AVW206 Interfaces in the network. Valid entries are 0-6. 27