Invacare ROZE Owners Manual - Page 25
Adjusting the Knee Pad Height,
View all Invacare ROZE manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 25 highlights
SECTION 7-MAINTENANCE NOTE: The actuator lead wires will be visible after the large base cover is removed. 6. Remove the socket head screw and washer that attach the leg actuator to the base. 7. At the other end of the actuator, remove the socket head screw and nut that attach the leg actuator to the leg bracket. 8. If necessary, repeat STEPS 5‐7 on the opposite side of the base to remove the other leg actuator. 9. To replace the actuator(s), reverse STEPS 5‐7. Caster BOTTOM VIEW OF PATIENT LIFT Pan head screw (eight screws attach the large base cover) Leg Large Base Cover Screw (eight screws attach each small base cover) Small Base Cover Mast Handles FIGURE 7.2 Replacing the Leg Actuators Adjusting the Knee Pad Height NOTE: For this procedure, refer to FIGURE 7.3. NOTE: When adjusted properly, the patient's knee will make contact at the center of the knee pad. 1. While supporting the knee pad, disengage the quick release knob on the side of the knee pad mounting bracket. 2. Choose a height adjustment that will be comfortable for the patient and will provide the necessary support. Engage the quick release knob into position until the knee pad is secure in place. Knee Pad Quick Release Knob CLOSE Part No. 1150703 FIGURE 7.3 Adjusting the Knee Pad Height 25 Roze™ Stand Up Patient Lift