Invacare ROZE Owners Manual - Page 21

Troubleshooting, SYMPTOMS, FAULTS, SOLUTION

Page 21 highlights

SECTION 6-TROUBLESHOOTING SYMPTOMS Noisy or dry sound from pivots. Electric actuator fails to lift when button is pressed. FAULTS Needs lubrication. Hand-control or actuator connector loose. Battery low. RED emergency stop button pressed IN. SECTION 6-TROUBLESHOOTING SOLUTION Refer to Lubricating the Lift on page 23. Charge batteries. Refer to Primary Emergency Release on page 13. Rotate RED emergency stop button CLOCKWISE until it pops out. Battery not connected properly to control box. The connecting terminals are damaged. Electric actuator in need of service or load is too high. Reconnect the battery to the control box. Refer to Primary Emergency Release on page 13. Replace the battery pack. Refer to Primary Emergency Release on page 13. Refer to Replacing the Electric Actuator on page 23. Unusual noise from lift arm actuator. Actuator is worn or damaged Refer to Replacing the Electric Actuator on or spindle is bent. page 23. Lift arms will not lower in uppermost position. Lift arms require a minimum weight load to lower from the uppermost position. Pull down slightly on the lift arms. Legs don't open and close properly. Leg actuators may be worn or Refer to Replacing the Leg Actuators on damaged or spindle is bent. page 24. NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare. Part No. 1150703 21 Roze™ Stand Up Patient Lift

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52

SECTION 6—TROUBLESHOOTING
Part No. 1150703
21
Roze™ Stand Up Patient Lift
SECTION 6—TROUBLESHOOTING
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare
.
SYMPTOMS
FAULTS
SOLUTION
Noisy or dry sound from pivots.
Needs lubrication.
Refer to Lubricating the Lift
on page 23.
Electric actuator fails to lift when button is
pressed.
Hand-control or actuator
connector loose.
Battery low.
RED emergency stop button
pressed IN.
Battery not connected properly
to control box.
The connecting terminals are
damaged.
Electric actuator in need of ser-
vice or load is too high.
Charge batteries. Refer to Primary Emer-
gency Release
on page 13.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the control box.
Refer to Primary Emergency Release
on
page 13.
Replace the battery pack. Refer to Primary
Emergency Release
on page 13.
Refer to Replacing the Electric Actuator
on
page 23.
Unusual noise from lift arm actuator.
Actuator is worn or damaged
or spindle is bent.
Refer to Replacing the Electric Actuator
on
page 23.
Lift arms will not lower in uppermost position.
Lift arms require a minimum
weight load to lower from the
uppermost position.
Pull down slightly on the lift arms.
Legs don’t open and close properly.
Leg actuators may be worn or
damaged or spindle is bent.
Refer to Replacing the Leg Actuators
on
page 24.