Troubleshooting, SYMPTOMS, FAULTS, SOLUTION
View all Invacare ROZE manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
SECTION 6-TROUBLESHOOTING SYMPTOMS Noisy or dry sound from pivots. Electric actuator fails to lift when button is pressed. FAULTS Needs lubrication. Hand-control or actuator connector loose. Battery low. RED emergency stop button pressed IN. SECTION 6-TROUBLESHOOTING SOLUTION Refer to Lubricating the Lift on page 23. Charge batteries. Refer to Primary Emergency Release on page 13. Rotate RED emergency stop button CLOCKWISE until it pops out. Battery not connected properly to control box. The connecting terminals are damaged. Electric actuator in need of service or load is too high. Reconnect the battery to the control box. Refer to Primary Emergency Release on page 13. Replace the battery pack. Refer to Primary Emergency Release on page 13. Refer to Replacing the Electric Actuator on page 23. Unusual noise from lift arm actuator. Actuator is worn or damaged Refer to Replacing the Electric Actuator on or spindle is bent. page 23. Lift arms will not lower in uppermost position. Lift arms require a minimum weight load to lower from the uppermost position. Pull down slightly on the lift arms. Legs don't open and close properly. Leg actuators may be worn or Refer to Replacing the Leg Actuators on damaged or spindle is bent. page 24. NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare. Part No. 1150703 21 Roze™ Stand Up Patient Lift