Invacare R51 Owners Manual - Page 63

Replacing the Battery Box Retaining Strap

Page 63 highlights

SECTION 12-BATTERIES Replacing the Battery Box Retaining Strap NOTE: For this procedure, refer to FIGURE 12.10 on page 64. ƽ WARNING The battery box retaining strap MUST be fastened securely and adjusted properly to hold battery boxes in place before using the wheelchair. 1. Remove the battery boxes from the wheelchair. Refer to Installing/Removing the Battery Boxes on page 53. 2. Remove the battery tray from the wheelchair. Refer to Installing/Removing the Battery Tray on page 54. ƽ WARNING The retaining strap MUST be fed through the adjustable end of the battery box retaining strap buckle as shown in FIGURE 12.10. NOTE: Observe the position and orientation of the battery box retaining strap as it passes through the slots of the battery tray and the adjustable end of the battery box retraining strap buckle. 3. Feed the existing battery box retaining strap through the adjustable end of the battery box retaining strap buckle. 4. Feed the existing battery box retaining strap through the slots in the battery box tray. 5. Install the NEW battery box retaining strap by reversing STEPS 3 and 4 as shown in FIGURE 12.10. 6. Reinstall the battery tray from the wheelchair. Refer to Installing/Removing the Battery Tray on page 54. 7. Reinstall the battery boxes from the wheelchair. Refer to Installing/Removing the Battery Boxes on page 53. Part No. 1106645 63 R51™and R51LXP

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76

SECTION 12—BATTERIES
Part No. 1106645
63
R51™and R51LXP
Replacing the Battery Box Retaining Strap
NOTE: For this procedure, refer to FIGURE 12.10 on page 64.
±
WARNING
The battery box retaining strap MUST be fastened securely and adjusted properly
to hold battery boxes in place before using the wheelchair.
1.
Remove the battery boxes from the wheelchair. Refer to Installing/Removing the
Battery Boxes
on page 53.
2.
Remove the battery tray from the wheelchair. Refer to Installing/Removing the Battery
Tray
on page 54.
±
WARNING
The retaining strap MUST be fed through the adjustable end of the battery box
retaining strap buckle as shown in FIGURE 12.10.
NOTE: Observe the position and orientation of the battery box retaining strap as it passes through
the slots of the battery tray and the adjustable end of the battery box retraining strap buckle.
3.
Feed the existing battery box retaining strap through the adjustable end of the battery
box retaining strap buckle.
4.
Feed the existing battery box retaining strap through the slots in the battery box tray.
5.
Install the NEW battery box retaining strap by reversing STEPS 3 and 4 as shown in
FIGURE 12.10.
6.
Reinstall the battery tray from the wheelchair. Refer to Installing/Removing the
Battery Tray
on page 54.
7.
Reinstall the battery boxes from the wheelchair. Refer to Installing/Removing the
Battery Boxes