Invacare TDXSP2 Owners Manual 2 - Page 48
Adjusting Multi-Axis Headrest Hardware, Installing, Adjusting Depth and Angle
View all Invacare TDXSP2 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 48 highlights
Invacare® Ultra Low Maxx by Motion Concepts 1. Loosen screws and clamping levers of dual link adjustment assembly A and screws of upper and lower rotational pivots B. 2. Set headrest to desired position. 3. Re-tighten screws and clamping levers. 4.11.3 Adjusting Multi-Axis Headrest Hardware Installing • 2.5 mm Allen key • 4 mm Allen key • 10 mm wrench 1. Using hardware provided, align and install headrest clamp assembly into existing mounting holes in backrest pan A. 2. Install headrest pad (not shown) to headrest rod using mounting hardware provided. The headrest pad can be adjusted to any desired angle via the pivot ball at the end of the headrest rod by loosening and tightening the mounting hardware. 3. Adjust overall height of headrest pad to desired position. Tighten knob C. For proper set-up headrest should be adjusted to user's ear height. 4. Once final height position is set, adjust D-Ring B so that it rests flush with top of clamp assembly (to prevent slipping). Adjusting Depth and Angle The headrest and horizontal rod can be further adjusted for depth and angle via the triangular multi-offset bracket. • 4 mm Allen key (5/32") Fig. 4-40 48 Fig. 4-41 60101832-A