HP Latex R2000 Site Preparation Guide - Page 11

Power distribution unit (PDU), Power specifications

Page 11 highlights

NOTE: The HP Internal Print Server power can be obtained by making a connection inside the electrical cabinet. 2. Power Distribution Unit (PDU) including single-phase branch circuit breaker (optional) 3. Power Distribution Unit (PDU) including three-phase branch circuit breaker depending on the power configuration NOTE: Remember that you are required to follow the local laws, regulations, and standards that apply to the electrical installation of your printer. NOTE: The printer is not supplied with any power cable. Power distribution unit (PDU) The PDU must be rated to meet the power requirements of the printer, and should be in accordance with the Electrical Code requirements of the local jurisdiction of the country where the equipment is installed. Power specifications Configuration 1: 380-415 V line-to-line three-phase configuration Three-phase specifications Number of power wires 5 (L1/L2/L3/N/PE) Input voltage (line to line) 380-415 V~ (±10%) Input frequency 50/60 Hz Maximum load current (per phase) 35 A Branch circuit breaker specifications Three-phase 4 poles, 40/50 A AC three-phase power cable specifications Configuration 5 wires, L1/L2/L3/N/PE Wire Strained Cu32 minimum, 10 mm² or 8 AWG Terminals Lines, ferrule terminals, PE, M8 ring terminal External diameter range 22.0-33.0 mm ENWW System operation requirements 7

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

NOTE:
The HP Internal Print Server power can be obtained by making a connection inside the electrical
cabinet.
2.
Power Distribution Unit (PDU) including single-phase branch circuit breaker (optional)
3.
Power Distribution Unit (PDU) including three-phase branch circuit breaker depending on the power
configuration
NOTE:
Remember that you are required to follow the local laws, regulations, and standards that apply to the
electrical installation of your printer.
NOTE:
The printer is not supplied with any power cable.
Power distribution unit (PDU)
The PDU must be rated to meet the power requirements of the printer, and should be in accordance with the
Electrical Code requirements of the local jurisdiction of the country where the equipment is installed.
Power specifications
Configuration 1: 380–415 V line-to-line three-phase configuration
Three-phase specifications
Number of power wires
5 (L1/L2/L3/N/PE)
Input voltage (line to line)
380–415 V~ (±10%)
Input frequency
50/60 Hz
Maximum load current (per phase)
35 A
Branch circuit breaker specifications
Three-phase
4 poles, 40/50 A
AC three-phase power cable specifications
Configuration
5 wires, L1/L2/L3/N/PE
Wire
Strained Cu32 minimum, 10 mm² or 8 AWG
Terminals
Lines, ferrule terminals, PE, M8 ring terminal
External diameter range
22.0–33.0 mm
ENWW
System operation requirements
7