Invacare RPL600-2 Owners Manual - Page 38

Warning, Caution

Page 38 highlights

10 MAINTENANCE 10.7 Replacing the Swivel Bar ƽ WARNING After the first year of use, the hooks of the swivel bar and mounting brackets of the boom should be inspected every six months to determine the extent of wear. If these parts become worn, replacement must be made. 1. The swivel bar comes attached to the boom with bolt, nut, washers, and pinch guard.  Washers should be placed against both sides of the boom mounting bracket to eliminate damage. The swivel bar has three hooks on both ends to attach the sling that supports the patient during lift. 2. Remove existing hardware and replace the swivel bar. ƽ CAUTION DO NOT overtighten the nut and bolt. This damages the mounting bracket. 450 lbs Swivel Bar Electric Portable Patient Lift 38 600 lbs Swivel Bar Part No. 1145810

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52

10 MAINTENANCE
Electric Portable Patient Lift
38
Part No. 1145810
10.7
Replacing the Swivel Bar
1.
The swivel bar comes attached to the boom with bolt, nut, washers, and pinch guard.
2.
Remove existing hardware and replace the swivel bar.
±
WARNING
After the first year of use, the hooks of the swivel bar and mounting brackets of the boom should be inspected every six months to
determine the extent of wear. If these parts become worn, replacement must be made.
Washers should be placed against both sides of the boom mounting bracket to eliminate damage.
The swivel bar has three hooks on both ends to attach the sling that supports the patient during lift.
±
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
600 lbs Swivel Bar
450 lbs Swivel Bar