Invacare RPL600-2 Owners Manual - Page 37

Checking and Tightening Mast Pivot Bolt

Page 37 highlights

10.6 Checking and Tightening Mast Pivot Bolt 1. Lift up the back of the rubber boot and slide it off the mast along the boom. 2. Check that the bolt is through the bracket and the locknut is tight and secure. 3. If needed, do one or more of the following: • Tighten locknut and back-off the locknut 1/8 of a turn. • Replace the locknut. 4. Reposition the rubber boot. Washer Bolt 10 MAINTENANCE Rubber Boot Boom Spacer (if equipped) Mast Locknut Bracket Part No. 1145810 37 Electric Portable Patient Lift

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52

10 MAINTENANCE
Part No. 1145810
37
Electric Portable Patient Lift
10.6
Checking and Tightening Mast Pivot Bolt
1.
Lift up the back of the rubber boot and slide it off the mast along
the boom.
2.
Check that the bolt is through the bracket and the locknut is tight
and secure.
3.
If needed, do one or more of the following:
Tighten locknut and back-off the locknut 1/8 of a turn.
Replace the locknut.
4.
Reposition the rubber boot.
Mast
Spacer
(if equipped)
Locknut
Boom
Bolt
Washer
Rubber Boot
Bracket