9 TROUBLESHOOTING 9 Troubleshooting SYMPTOMS Patient lift feels loose. Casters/Brakes noisy or stiff. Noisy or dry sound from pivots. Oil leaking from hydraulics. FAULTS Mast/Base joint loose. Tie - Rods are loose. Fluff or debris in bearings. Needs lubrication. Hydraulic pump in need of replacement. Manual / Hydraulic pump fails to lift when pumped. Control Valve not fully closed. Manual / Hydraulic pump in need of replacement. SOLUTION Refer to Assembly on page 15. Refer to Maintaining the Base Adjustment on page 48. Refer to Replacing Casters/Forks on page 48. Refer to Lubricating the Lift on page 43. Refer to Replacing a Manual/Hydraulic Pump on page 44 and Replacing the Electric Actuator on page 45. Contact your Dealer. Close Control Valve. Refer to Replacing a Manual/Hydraulic Pump on page 44 and Replacing the Electric Actuator on page 45. Contact your Dealer. Check connections. Manual/Electric Portable Patient Lift 38 Part No. 1078987

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60

9 TROUBLESHOOTING
Manual/Electric Portable Patient Lift
38
Part No. 1078987
9
Troubleshooting
SYMPTOMS
FAULTS
SOLUTION
Patient lift feels loose.
Mast/Base joint loose.
Tie - Rods are loose.
Refer to Assembly
on page 15.
Refer to Maintaining the Base Adjustment
on page 48.
Casters/Brakes noisy or stiff.
Fluff or debris in bearings.
Refer to Replacing Casters/Forks
on page 48.
Noisy or dry sound from pivots.
Needs lubrication.
Refer to Lubricating the Lift
on page 43.
Oil leaking from hydraulics.
Hydraulic pump in need of replacement.
Refer to Replacing a Manual/Hydraulic Pump
on page 44 and
Replacing the Electric Actuator
on page 45. Contact your Dealer.
Manual / Hydraulic pump fails to
lift when pumped.
Control Valve not fully closed.
Manual / Hydraulic pump in need of
replacement.
Close Control Valve.
Refer to Replacing a Manual/Hydraulic Pump
on page 44 and
Replacing the Electric Actuator
on page 45. Contact your Dealer.