Invacare IOH200 Owners Manual - Page 32

Technical Data

Page 32 highlights

Invacare® HomeFill® Oxygen System 8 Technical Data 8.1 Technical Description The HomeFill Oxygen System (IOH200) is an electromechanical prescription system for use in the home. Its intended function and use is to supply pressurized oxygen to fill Invacare proprietary oxygen cylinders for ambulatory use while the patient receives therapeutic oxygen from an oxygen concentrator. The IOH200 HomeFill Compressor is designed for use with oxygen concentrators that meet the input specifications for pressure, flow, and oxygen purity. The HomeFill Oxygen System consists of the IOH200 oxygen compressor, connection tubing, and the Invacare supplied portable oxygen cylinders with interlocking fill fitting. The IOH200 compressor consists of a multistage compressor, internal reservoir, control electronics with an integral oxygen concentration sensor, interlocking fill fitting, inlet fitting, and drive motor. The compressor operation is controlled by the electronics assembly. The motor/compressor is allowed to fill a cylinder only when the oxygen concentration of the inlet gas stream is greater than 90%. The motor/compressor fill operation is completed when the gas in a cylinder reaches the maximum fill pressure of 2000 PSI (13786 kPa). The IOH200 oxygen compressor can fill cylinders for portable use by ambulatory patients. The Invacare supplied portable gas cylinders have an interlocking fill fitting that mates only with the Invacare IOH200 compressor assembly. 32 8.2 Specifications Regulatory Listings Regulatory Listings: Double Insulated Product ETL Certified as Complying with UL60601-1, First Edition ETL Certified Complying with CAN/CSA C22.2 No. 601-1 M:90 (Canada) EMC IEC 60601-1-2 Input Specifications Input Pressure Required: Input Flow Required: Input Oxygen Purity Required: 5 - 25 psi 2 L/min >90% O2 General Specifications Electrical Requirements: 120 VAC ± 10% (108 Voltage Input: VAC/132 VAC), 60 Hz Rated Current Input: 2.0 AMP Operating Temperature: 50 to 95° F (10 to 35° C) @ 20 to 60% non-condensing humidity Width: 20 ¼ inches 1100873-N

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

Invacare® HomeFill® Oxygen System
8
Technical Data
8.1
Technical Description
The HomeFill Oxygen System (IOH200) is an
electromechanical prescription system for use in the home.
Its intended function and use is to supply pressurized oxygen
to fill Invacare proprietary oxygen cylinders for ambulatory
use while the patient receives therapeutic oxygen from an
oxygen concentrator. The IOH200 HomeFill Compressor is
designed for use with oxygen concentrators that meet the
input specifications for pressure, flow, and oxygen purity.
The HomeFill Oxygen System consists of the IOH200 oxygen
compressor, connection tubing, and the Invacare supplied
portable oxygen cylinders with interlocking fill fitting.
The IOH200 compressor consists of a multistage compressor,
internal reservoir, control electronics with an integral oxygen
concentration sensor, interlocking fill fitting, inlet fitting, and
drive motor. The compressor operation is controlled by the
electronics assembly. The motor/compressor is allowed to
fill a cylinder only when the oxygen concentration of the
inlet gas stream is greater than 90%. The motor/compressor
fill operation is completed when the gas in a cylinder
reaches the maximum fill pressure of 2000 PSI (13786 kPa).
The IOH200 oxygen compressor can fill cylinders for portable
use by ambulatory patients. The Invacare supplied portable
gas cylinders have an interlocking fill fitting that mates only
with the Invacare IOH200 compressor assembly.
8.2
Specifications
Regulatory Listings
Regulatory Listings:
Double Insulated Product
ETL Certified as Complying
with UL60601-1, First
Edition
ETL Certified Complying
with CAN/CSA C22.2 No.
601-1 M:90 (Canada)
EMC IEC 60601-1-2
Input Specifications
Input Pressure Required:
5 - 25 psi
Input Flow Required:
2 L/min
Input Oxygen Purity
Required:
>90% O
2
General Specifications
Electrical Requirements:
Voltage Input:
120 VAC ± 10% (108
VAC/132 VAC), 60 Hz
Rated Current Input:
2.0 AMP
Operating Temperature:
50 to 95° F (10 to 35° C) @
20 to 60% non-condensing
humidity
Width:
20 ¼ inches
32
1100873-N