Invacare IOH200 Owners Manual

Invacare IOH200 Manual

Invacare IOH200 manual content summary:

  • Invacare IOH200 | Owners Manual - Page 1
    Invacare® HomeFill® Oxygen System Model IOH200 en Compressor User Manual This manual MUST be given to the user of the product. BEFORE using this product, this manual MUST be read and saved for future reference.
  • Invacare IOH200 | Owners Manual - Page 2
    the U.S.A. 3-in-1 oil is a registered trademark of American Home Products Corporation. WD-40 is a registered trademark of the WD-40 Company. Updated versions of this manual are available on www.invacare.com.
  • Invacare IOH200 | Owners Manual - Page 3
    Info 22 5.2 Wear and Tear 22 5.3 Service Life 22 5.4 Cleaning the Compressor Filter 22 5.5 Cleaning the Cabinet 23 5.6 Replacing the Fuse 24 6 After Use 25 6.1 Disposal 25 7 Troubleshooting 26 7.1 Indicator Lights 26 7.2 Troubleshooting 28 8 Technical Data 32 8.1 Technical Description
  • Invacare IOH200 | Owners Manual - Page 4
    Invacare® HomeFill® Oxygen System 1 General 1.1 Symbols Symbols in the Manual Signal words are used in this manual trouble-free use. 4 Symbols on the Product The following symbols are found on the compressor, compressor packaging or the oxygen cylinder. Read the Manual (Compressor) Read the Manual
  • Invacare IOH200 | Owners Manual - Page 5
    Refer to 6.1 Disposal, page 25. 1.2 Intended Use The Invacare HomeFill compressor was designed to be used only with oxygen concentrators that under adult supervision only after reading and understanding the instructions and warnings of this user manual. WARNING! - The use of this device is limited
  • Invacare IOH200 | Owners Manual - Page 6
    understanding these instructions and any additional instructional material such as user manuals, service manuals or instruction sheets supplied technician or Invacare for repair. - In order to ensure the safe installation, assembly and operation of the HomeFill Compressor these instructions MUST be
  • Invacare IOH200 | Owners Manual - Page 7
    come into contact with the compressor coupling or cylinder fittings. Invacare suggests you wash your hands prior to using the HomeFill servicing to qualified personnel. WARNING! Risk of Serious Injury or Damage Use of non-Invacare accessories may result in serious injury or damage. - Invacare
  • Invacare IOH200 | Owners Manual - Page 8
    the cylinder outdoors and notify your Home Health Care Provider or Service Representative of this condition. - Children should ALWAYS be supervised around the HomeFill unit. When used with the HomeFill table or similar elevated support surfaces, a child could pull over any such table causing the
  • Invacare IOH200 | Owners Manual - Page 9
    Safety 2.2 Label Locations B C D E A 1100873-N 9
  • Invacare IOH200 | Owners Manual - Page 10
    Invacare recommends leaving a full cylinder of oxygen with the patient after setting up the HomeFill compressor. Before leaving a HomeFill Compressor with a user, the following checklist MUST be completed: q Make sure the pressure test on cylinder has not expired. q Instruct in this manual. This
  • Invacare IOH200 | Owners Manual - Page 11
    Safety device operation stops, then the compressor is causing the interference. In such rare cases, interference may • Connect either line powered device to a different electrical power circuit. Contact Invacare for any additional EMC information or EMC tables for the device environment. 1100873-N
  • Invacare IOH200 | Owners Manual - Page 12
    Invacare® HomeFill® Oxygen System 3 Setup 3.1 Unpacking 12 PART NUMBER 1102311 1102315 1105177 1112949 1100873 DESCRIPTION Compressor Assembly Packaging Assembly Supply Line Assembly Power Cord User Manual (not shown) 1. Check for any obvious damage to the carton or its contents. If damage is
  • Invacare IOH200 | Owners Manual - Page 13
    NOT SMOKE while using this device. - DO NOT use near OPEN FLAME or IGNITION SOURCES. - DO NOT use any lubricants on the compressor unless recommended by Invacare. - NO SMOKING signs should be prominently displayed. - Avoid creation of any spark near oxygen equipment. This includes sparks from static
  • Invacare IOH200 | Owners Manual - Page 14
    Invacare® HomeFill® Oxygen System 3.5 Installing the Power Cord 1. There are two methods to safely transport the compressor: • One Hand - Grasp the hand grip A at the bottom rear of the compressor. • Two Hands - Grasp the hand grips B on the sides of compressor. 14 This product may come with a
  • Invacare IOH200 | Owners Manual - Page 15
    -Invacare Concentrators 1. Refer to the oxygen concentrator user manual outlet gas specifications to verify compatibility with the HomeFill compressor input gas requirements listed in 8.2 Specifications, page 32. 2. Refer to the oxygen concentrator user manual for connecting instructions. 1100873
  • Invacare IOH200 | Owners Manual - Page 16
    Invacare® HomeFill® Oxygen System 4 Usage 4.1 Compressor Operation Checklist Each time the HomeFill Compressor is used to fill a cylinder, complete the following checklist: q Ensure the concentrator has been On for at least thirty minutes. Refer to the concentrator User Manual. q Perform the prefill
  • Invacare IOH200 | Owners Manual - Page 17
    inside the valve • Signs of excessive heat or fire damage Usage 4.3 Connecting the Cylinder to the Compressor WARNING! - NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. WARNING! - DO NOT drop oxygen cylinders. Use
  • Invacare IOH200 | Owners Manual - Page 18
    Invacare® HomeFill® Oxygen System Fillport covers not shown for clarity. 1. Examine the cylinder gauge. If the cylinder pressure is less than 1500 psig, proceed to STEP 2. For location of the cylinder gauge, refer to the cylinder's manual Position the cylinder in the compressor cradle G. 7. Align the
  • Invacare IOH200 | Owners Manual - Page 19
    cradle. WARNING! - NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. 1. Press the compressor power switch A to the Off (O) position. 2. Grasp the cylinder B in the area behind the cylinder fillport C. 1100873
  • Invacare IOH200 | Owners Manual - Page 20
    to one of the following settings. Refer to the concentrator User Manual. • 3 L/min or less for 5L models • 5 L/min or less for 10L models Non-Invacare concentrators may require lower output flow rates to the patient in order to provide ≥ 90% oxygen levels to the HomeFill compressor. 1100873-N
  • Invacare IOH200 | Owners Manual - Page 21
    oxygen level from the concentrator has not yet reached or drops below 90%. The compressor will begin or resume filling when the oxygen level reaches 90%. If light does not go Off within 10 minutes, refer to 7.2 Troubleshooting, page 28. If the ATTENTION (RED) light and audible alarm F are on refer
  • Invacare IOH200 | Owners Manual - Page 22
    of this product is 5 years or 5000 hours of operation when used in accordance with the safety instructions, maintenance intervals and correct use, stated in this manual. The effective service life can vary according to frequency and intensity of use. Maintenance should be done at least every week
  • Invacare IOH200 | Owners Manual - Page 23
    1. Unplug the compressor A. Grasp the plug not the cord. 2. Remove the filter B. 3. Clean the filter with a of Electric Shock This unit contains moving and high pressure components. There are no user serviceable parts inside the cover. To avoid electric shock, injury or death: - DO NOT remove the cover
  • Invacare IOH200 | Owners Manual - Page 24
    Invacare® HomeFill® Oxygen System 5.6 Replacing the Fuse 24 1. Unplug the power cord from the electrical outlet. 2. Remove the power cord from the power inlet A on the compressor base C. 3. Pull out the fuse drawer B at the bottom of the power inlet. 4. Using a fuse puller or a small flat blade
  • Invacare IOH200 | Owners Manual - Page 25
    be handled in accordance with the individual product marking for disposal. Please be environmentally responsible and recycle this product at its end of life. Contact Invacare or your service provider for product recycling information. 1100873-N After Use 25
  • Invacare IOH200 | Owners Manual - Page 26
    Invacare® HomeFill® Oxygen System 7 Troubleshooting 7.1 Indicator Lights Indicator Light Color None O2 BELOW NORMAL - YELLOW FILLING - GREEN Compressor Status Effect Compressor is Off. O2 flowing to user if concentrator is on. Compressor is On and warming up (0-3 Cylinder filling not started.
  • Invacare IOH200 | Owners Manual - Page 27
    Indicator Light Color FULL - GREEN ATTENTION - RED (Audible Alarm) Troubleshooting Compressor Status Effect Compressor is On. Cylinder is finished filling (full) -remove cylinder (FULL light is On in >10 mins.) or Cylinder is not attached properly - check connection (FULL
  • Invacare IOH200 | Owners Manual - Page 28
    Invacare® HomeFill® Oxygen System 7.2 Troubleshooting Problem No indicator lights on when compressor turned On. RED Light On 28 Solution 1. Turn compressor Off. 2. Check that power cord is properly plugged into compressor and wall outlet. 3. Turn the compressor On. 4. Clean air filters. Refer to
  • Invacare IOH200 | Owners Manual - Page 29
    does not appear, proceed to step 2. 2. The concentrator may need service, contact your home health care provider or Invacare. 1. Ensure the concentrator has warmed up for at least 20 minutes. If it has not been 20 minutes, turn the compressor Off until the concentrator warms up. If the YELLOW light
  • Invacare IOH200 | Owners Manual - Page 30
    setting (refer to the concentrator User manual). If the GREEN filling light does not go Off, the compressor and/or concentrator may need service. Contact your home health care provider or Invacare. 1. Call home health care provider or Invacare for service. 1. Turn flow selector Off. 2. If hissing
  • Invacare IOH200 | Owners Manual - Page 31
    Refer to the cylinder's manual for proper fill times. Filling times are approximate and may vary on environmental conditions. Troubleshooting 1100873-N 31
  • Invacare IOH200 | Owners Manual - Page 32
    . Its intended function and use is to supply pressurized oxygen to fill Invacare proprietary oxygen cylinders for ambulatory use while the patient receives therapeutic oxygen from an oxygen concentrator. The IOH200 HomeFill Compressor is designed for use with oxygen concentrators that meet the input
  • Invacare IOH200 | Owners Manual - Page 33
    Height: Depth: Weight: Shipping Weight: No AP/APG: Electrical: Placement: Time of Operation: Recommended Storage and Shipping Temperature: Environment: Minimum Operating Time: Altitude: 15 inches 16 inches 33 lbs 39 lbs Not suitable for use in the presence of a flammable anesthetic mixture. No
  • Invacare IOH200 | Owners Manual - Page 34
    state to state. Invacare warrants the HomeFill Compressor to be free from defects in Invacare Service Department or Invacare dealers. It does not apply to the labor performed by the purchaser or user. This warranty does not include normal wear and tear, failure to adhere to the product instructions
  • Invacare IOH200 | Owners Manual - Page 35
    cost, to return the unit, pack the unit in a manner to avoid shipping damage and to ship the unit to either Invacare's plant or service center as specified by Invacare in advance. Defective units must be returned for warranty inspection using the serial number as identification within thirty days of
  • Invacare IOH200 | Owners Manual - Page 36
    Invacare Corporation USA One Invacare Way Elyria, Ohio USA 44035 800-333-6900 Technical Services: 800-832-4707 www.invacare.com Manufacturer Invacare Corporation 2101 E. Lake Mary Blvd. Sanford, FL USA 32773 407-321-5630 1100873-N 2019-02-13 *1100873N* Making Life's Experiences Possible®
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

InvacareĀ® HomeFillĀ® Oxygen
System
Model IOH200
en
Compressor
User Manual
This manual MUST be given to the user of the product.
BEFORE using this product, this manual MUST be read and saved for
future reference.