SECTION 7-MAINTENANCE SECTION 7-MAINTENANCE Maintenance Safety Inspection Checklist ITEM INITIALLY INSTITUTIONAL INSPECT/ADJUST MONTHLY IN-HOME INSPECT EVERY SIX (6) MONTHS THE BASE Inspect for missing hardware. X X X Legs open/close with ease. X X X Inspect casters and axle bolts for tightness and wear. X X X Inspect casters for smooth swivel and roll. X X X Inspect wheels are clear of debris. X X X THE MAST Mast MUST be securely assembled to lift arms. X X X Inspect for bends or deflections. X X X THE LIFT ARMS AND LINKAGE Check all hardware and attachment points. X X X Inspect for bends or deflections. X X X Inspect bolted joints of lift arms for wear. X X X Inspect to ensure that the lift arms are centered between the base legs. X X X ELECTRIC ACTUATOR Check for wear or deterioration. (IF DAMAGED, CONTACT DEALER) X X X Cycle to ensure smooth quiet operation. X X X CLEANING Whenever necessary. X X X SLINGS CHECK ALL SLING ATTACHMENTS each time it is used to ensure proper connection and patient safety. X X X Inspect sling material for wear. X X X Inspect straps for wear. X X X NOTE: Follow the maintenance procedures described in this manual to keep your patient lift in continuous service. The Invacare Patient Lift is designed to provide a maximum of safe, efficient and satisfactory service with minimum care and maintenance. All parts of the Invacare Lift are made of the best grades of steel, but metal to metal contact will wear after considerable use. There is no adjustment or maintenance of the casters, other than cleaning, lubrication and checking axle and swivel bolts for tightness. Remove all debris, etc. from the wheel and swivel bearings. If any parts are worn, replace these parts IMMEDIATELY. If you question the safety of any part of the lift, contact your Dealer IMMEDIATELY. Cleaning the Sling and the Lift The sling should be washed regularly in water temperature not exceeding 180°F (82°C) and a biological solution. A soft cloth, dampened with water and a small amount of mild detergent, is all that is needed to clean the patient lift. The lift can be cleaned with non‐abrasive cleaners. Detecting Wear and Damage It is important to inspect all stressed parts, such as slings, lifting arm and any pivot for slings for signs of cracking, fraying, deformation or deterioration. Replace any defective parts IMMEDIATELY and ensure that the lift is not used until repairs are made. Part No. 1078984 21 Stand Up Patient Lift

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

SECTION 7—MAINTENANCE
Part No. 1078984
21
Stand Up Patient Lift
SECTION 7—MAINTENANCE
Maintenance Safety Inspection Checklist
NOTE: Follow the maintenance procedures described in this manual to keep your patient lift in continuous service.
The Invacare Patient Lift is designed to provide a maximum of safe, efficient and satisfactory service with minimum care and
maintenance.
All parts of the Invacare Lift are made of the best grades of steel, but metal to metal contact will wear after considerable use.
There is no adjustment or maintenance of the casters, other than cleaning, lubrication and checking axle and swivel bolts for
tightness. Remove all debris, etc. from the wheel and swivel bearings. If any parts are worn, replace these parts IMMEDIATELY.
If you question the safety of any part of the lift, contact your Dealer IMMEDIATELY.
Cleaning the Sling and the Lift
The sling should be washed regularly in water temperature not exceeding 180°F (82°C) and a biological solution. A soft
cloth, dampened with water and a small amount of mild detergent, is all that is needed to clean the patient lift. The lift can
be cleaned with non
abrasive cleaners.
Detecting Wear and Damage
It is important to inspect all stressed parts, such as slings, lifting arm and any pivot for slings for signs of cracking, fraying,
deformation or deterioration. Replace any defective parts IMMEDIATELY and ensure that the lift is not used until repairs
are made.
ITEM
INITIALLY
INSTITUTIONAL
INSPECT/ADJUST
MONTHLY
IN-HOME INSPECT
EVERY SIX (6)
MONTHS
THE BASE
Inspect for missing hardware.
Legs open/close with ease.
Inspect casters and axle bolts for tightness and wear.
Inspect casters for smooth swivel and roll.
Inspect wheels are clear of debris.
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
THE MAST
Mast MUST be securely assembled to lift arms.
Inspect for bends or deflections.
X
X
X
X
X
X
THE LIFT ARMS AND LINKAGE
Check all hardware and attachment points.
Inspect for bends or deflections.
Inspect bolted joints of lift arms for wear.
Inspect to ensure that the lift arms are centered
between the base legs.
X
X
X
X
X
X
X
X
X
X
X
X
ELECTRIC ACTUATOR
Check for wear or deterioration.
(IF DAMAGED, CONTACT DEALER)
Cycle to ensure smooth quiet operation.
X
X
X
X
X
X
CLEANING
Whenever necessary.
X
X
X
SLINGS
CHECK ALL SLING ATTACHMENTS each time it is
used to ensure proper connection and patient safety.
Inspect sling material for wear.
Inspect straps for wear.
X
X
X
X
X
X
X
X
X