Invacare RPL450-2 Owners Manual - Page 36

Detecting Wear and Damage, Cleaning the Sling and the Lift, Replacing the Boom Actuator

Page 36 highlights

10 MAINTENANCE 10.3 Detecting Wear and Damage It is important to inspect all stressed parts, such as slings, spreader bar and any pivot for slings for signs of cracking, fraying, deformation or deterioration. Replace any defective parts immediately and ensure that the lift is not used until repairs are made. 10.4 Cleaning the Sling and the Lift The sling should be regularly washed in water, temperature not to exceed 180°F (82°C) and a biocidal (anti-biological) solution. A soft cloth, dampened with water and a small amount of mild detergent, is all that is needed to clean the patient lift. The lift can be cleaned with non-abrasive cleaners. 10.5 Replacing the Boom Actuator ƽ CAUTION DO NOT overtighten the nut and bolt. This damages the mounting bracket. Boom 1. Remove the nut, washer and shoulder bolt that secure the electric actuator to the mast mounting bracket. 2. Rest the boom on your shoulder and remove the nut, bolt, plastic bushing and pinch guard from the boom mounting bracket. 3. Remove the boom actuator assembly. 4. Reverse steps for installation. Mast Pinch Guard Bushing Electric Portable Patient Lift 36 Part No. 1145810

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52

10 MAINTENANCE
Electric Portable Patient Lift
36
Part No. 1145810
10.3
Detecting Wear and Damage
It is important to inspect all stressed parts, such as slings, spreader bar and any pivot for slings for signs of cracking, fraying, deformation or
deterioration. Replace any defective parts immediately and ensure that the lift is not used until repairs are made.
10.4
Cleaning the Sling and the Lift
The sling should be regularly washed in water, temperature not to exceed 180°F (82°C) and a biocidal (anti-biological) solution. A soft cloth, dampened
with water and a small amount of mild detergent, is all that is needed to clean the patient lift. The lift can be cleaned with non-abrasive cleaners.
10.5
Replacing the Boom Actuator
1.
Remove the nut, washer and shoulder bolt that secure the electric
actuator to the mast mounting bracket.
2.
Rest the boom on your shoulder and remove the nut, bolt, plastic
bushing and pinch guard from the boom mounting bracket.
3.
Remove the boom actuator assembly.
4.
Reverse steps for installation.
±
CAUTION
DO NOT overtighten the nut and bolt. This damages the
mounting bracket.
Boom
Pinch
Guard
Bushing
Mast