9. Perform the following steps to clean the control unit and hose fittings: a. Wipe all controls, chassis and hose fittings with a quaternary disinfectant solution. b. Using a nylon brush, gently clean all crevices as they can harbor microorganisms. c. Air dry all treated surfaces. 10. Perform the following steps to clean the mattress components (air cushions, mattress base, etc.): a. Using a clean sponge or paper towel, wipe down the mattress components with a dilute detergent solution of quaternary cleaner disinfectant or other germicidal detergent solution. b. Wipe the mattress components with a clean dry cloth. 11. Test alarm function for power fail and low pressure. Maintenance 6.3 Changing the Filter Clean the filter every 5 months or whenever it is dirty. AB A 1148138-I 1. Remove the two screws A securing the filter cover B to the back of the control unit. 2. Remove the filter. 3. Wash the filter with water and a mild detergent. 4. Dry the filter completely. 5. Install the filter into the back of the control unit. 6. Secure the filter cover to the back of the control unit using the two screws. 33

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40

Maintenance
9.
Perform the following steps to clean the control unit
and hose fittings:
a. Wipe all controls, chassis and hose fittings with a
quaternary disinfectant solution.
b. Using a nylon brush, gently clean all crevices as they
can harbor microorganisms.
c. Air dry all treated surfaces.
10. Perform the following steps to clean the mattress
components (air cushions, mattress base, etc.):
a. Using a clean sponge or paper towel, wipe down the
mattress components with a dilute detergent solution
of quaternary cleaner disinfectant or other germicidal
detergent solution.
b. Wipe the mattress components with a clean dry
cloth.
11. Test alarm function for power fail and low pressure.
6.3
Changing the Filter
Clean the filter every 5 months or whenever it is
dirty.
1.
Remove the two screws
A
securing the filter cover
B
to the back of the control unit.
2.
Remove the filter.
3.
Wash the filter with water and a mild detergent.
4.
Dry the filter completely.
5.
Install the filter into the back of the control unit.
6.
Secure the filter cover to the back of the control unit
using the two screws.
1148138-I