Invacare IRC5P Owners Manual - Page 27
Troubleshooting Guide, Symptom, Probable Cause, Solution - irc5po2
View all Invacare IRC5P manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 27 highlights
SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Alarm: Continuous On RED light only. * Only applies to IRC5PO2. PROBABLE CAUSE: 1. Low oxygen purity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min or less.* 4. Unit overheating due to blocked air intake. 5. Insufficient power at outlet. 6. Internal repairs required. SOLUTION: 1. Clean or Replace filters. 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back On. 3. Check flowmeter is set to 1.0 L/min or more. 4a. Remove and clean cabinet filters. 4b. Move concentrator at least three inches from walls, draperies, and furniture. 5. DO NOT use extension cords. Move to another electrical outlet or circuit. 6. Call Service Provider. GREEN light with 1. Internal repairs YELLOW Light required. Flashing. 1. Call Service Provider. Part No 1143482 27 Perfecto2™Series