Invacare IRC5P Owners Manual

Invacare IRC5P Manual

Invacare IRC5P manual content summary:

  • Invacare IRC5P | Owners Manual - Page 1
    Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this product. User: BEFORE using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services
  • Invacare IRC5P | Owners Manual - Page 2
    READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER'S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS and cleaning. NOTE: Updated versions of this manual can be found at www.invacare.com. Perfecto2™Series 2 Part No 1143482
  • Invacare IRC5P | Owners Manual - Page 3
    Switch 18 Flowrate 19 Front Panel Indicators - Units with SensO2 20 Front Panel Indicators - Units without SensO2 21 SECTION 6-MAINTENANCE 23 Routine Maintenance 23 SECTION 7-TROUBLESHOOTING GUIDE ......... 26 LIMITED WARRANTY 31 Part No 1143482 3 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 4
    are used in this manual and apply to hazards room in which this product is located and away from where oxygen is being delivered. NO SMOKING signs should be prominently displayed. burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result
  • Invacare IRC5P | Owners Manual - Page 5
    practices to use or order the use of this device." Invacare recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition or be used as an oxygen supplement and is not considered life supporting or life sustaining. Part No 1143482 5 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 6
    LABEL LOCATION LABEL LOCATION Perfecto2™ FOR OPERATING INSTRUCTIONS or ALARMS (See Operator's Manual) DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where oxygen is being delivered. Textiles, oil and other
  • Invacare IRC5P | Owners Manual - Page 7
    concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless recommended by Invacare. Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction. Part No 1143482 7 Perfecto2
  • Invacare IRC5P | Owners Manual - Page 8
    . DO NOT move or relocate concentrator by pulling on the power cord. NEVER drop or insert any object or liquid into any opening. Invacare recommends that Crush‐Proof oxygen tubing be used with this product and not exceed 50 ft in length. There are no user serviceable parts. This does not include
  • Invacare IRC5P | Owners Manual - Page 9
    , or increase the separation between the equipment. • Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected. Part No 1143482 9 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 10
    ʹs manual, part number 1100873, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet Power Cord *Outlet Fitting fitting. For more information about the HomeFill, contact your Invacare dealer. Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 11
    Inspection Inspect/examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components. Storage 1. Store the repackaged oxygen concentrator in a dry area. 2. DO NOT place objects on top of repackaged concentrator. Part No 1143482 11 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 12
    *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). Perfecto2 and Perfecto2W Models
  • Invacare IRC5P | Owners Manual - Page 13
    7kPa is applied: 0.7 L/min Filters: Cabinet, Outlet HEPA and Compressor Perfecto2 Models 50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto2W Models 45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto2V Model 50°F - 95°F (10°C - 35°C) at 20-60% relative humidity Perfecto2 Models and Perfecto2V Model Exhaust: Oxygen
  • Invacare IRC5P | Owners Manual - Page 14
    (DO NOT pinch). 20 to 60% Up to 24 hours per day. -20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity. *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes). Perfecto2™Series 14 Part No 1143482
  • Invacare IRC5P | Owners Manual - Page 15
    oxygen delivery systems. Your provider will show you how to use your oxygen concentrator. He/She should be contacted with any questions or problems regarding your oxygen concentrator. This owner's manual will tell you about your concentrator place in a closet. Part No 1143482 15 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 16
    w/o Cap FIGURE 5.1 Filling the Humidifier 3. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door off (FIGURE 5.2). 4. Pull up and remove the humidifier bottle adapter (FIGURE 5.2). Perfecto2™Series 16 Part No 1143482
  • Invacare IRC5P | Owners Manual - Page 17
    to oxygen outlet connector on the concentrator. See Detail "B" in FIGURE 5.3. 8. Attach the cannula/patient supply tubing to the humidifier bottle outlet. See Detail "B" in FIGURE 5.3. 9. After assembly, ensure that oxygen is flowing through the cannula. Part No 1143482 17 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 18
    5-OPERATING INSTRUCTIONS DETAIL "A" DETAIL "B" Humidifier Bottle Oxygen Tubing Adapter Humidifier Bottle Humidifier Outlet Bottle Humidifier Bottle Oxygen Outlet Connector is functioning properly. On/Off (I/O) Power Switch FIGURE 5.4 Power Switch Perfecto2™Series 18 Part No 1143482
  • Invacare IRC5P | Owners Manual - Page 19
    SECTION 5-OPERATING INSTRUCTIONS Flowrate NOTE: For this procedure, refer to FIGURE 5.5 on page 20. 1. Turn the flowrate knob to the setting use of some accessories such as the pediatric flowstand and the HomeFill compressor will deactivate the Low Flow Alarm. Part No 1143482 19 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 20
    concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate. OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators - Units with SensO2 Perfecto2™Series 20 Part
  • Invacare IRC5P | Owners Manual - Page 21
    INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator qualified technician. Have a back-up supply of oxygen ready. RED Indicator Light Continuous Audible Alarm Sieve-GARD technician. Switch to a back-up supply of oxygen immediately. Front Panel Indicators - Units without SensO2
  • Invacare IRC5P | Owners Manual - Page 22
    SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min). ™Compressor Shutdown Call a qualified technician. Switch to a back-up supply of oxygen immediately. Perfecto2™Series 22 Part No 1143482
  • Invacare IRC5P | Owners Manual - Page 23
    back of the cabinet. 1. Remove the filter and clean at least once a week depending on environmental conditions. NOTE: Environmental conditions that may require more frequent cleaning of the filters include but are not limited to: high dust, air pollutants, etc. Part No 1143482 23 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 24
    before reinstallation. FIGURE 6.1 Cleaning the Cabinet Filter Cleaning the Cabinet Clean the cabinet with a mild household cleaner and non‐abrasive cloth or sponge. Cleaning the Humidifier NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are
  • Invacare IRC5P | Owners Manual - Page 25
    Maintenance Record Part No 1143482 25 SECTION 6-MAINTENANCE SERIAL NO. PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT
  • Invacare IRC5P | Owners Manual - Page 26
    repairs required. 1a. Remove and clean cabinet filters. 1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture. 2. DO NOT use extension cords. Move to another electrical outlet or circuit. 3. Call Service Provider. Alarm: Rapid Beep....Beep... Beep...Beep
  • Invacare IRC5P | Owners Manual - Page 27
    SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Alarm: Continuous On RED light only. * Only applies to IRC5PO2. PROBABLE CAUSE: 1. Low oxygen purity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min or less.* 4. Unit overheating due to
  • Invacare IRC5P | Owners Manual - Page 28
    NOTES NOTES Perfecto2™Series 28 Part No 1143482
  • Invacare IRC5P | Owners Manual - Page 29
    NOTES NOTES Part No 1143482 29 Perfecto2™Series
  • Invacare IRC5P | Owners Manual - Page 30
    NOTES NOTES Perfecto2™Series 30 Part No 1143482
  • Invacare IRC5P | Owners Manual - Page 31
    rights which may vary from state to state. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be free from items, such as filters, are excluded from this warranty. For warranty service, please contact Invacare Corporation's Service Department at the toll
  • Invacare IRC5P | Owners Manual - Page 32
    TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS. Invacare Corporation USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 Part No 1143482 www.invacare.com © 2009 Invacare Corporation. All rights reserved. Republication, duplication or modification
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

Dealer:
This manual MUST be given to the end
user of this product.
User:
BEFORE using this product, read this
manual and save for future reference.
User Manual
For more information regarding
Invacare products,
parts, and services,
please visit www.invacare.com
Perfecto
2
Series
Oxygen Concentrators without SensO
2
®,
Oxygen Concentrators with SensO
2
HomeFill
®
Compatible
Perfecto
2
with SensO
2
Model IRC5PO2
Model IRC5PO2W
Model IRC5PO2V
Perfecto
2
Model IRC5P
Model IRC5PW