Invacare MA65 Owners Manual

Invacare MA65 Manual

Invacare MA65 manual content summary:

  • Invacare MA65 | Owners Manual - Page 1
    microAIR® MA65 Series MA65, MA65B42, MA65B48, MA65RSR en Alternating Pressure Low Air Loss Mattress User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
  • Invacare MA65 | Owners Manual - Page 2
    or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. Making Life's Experiences Possible is a registered
  • Invacare MA65 | Owners Manual - Page 3
    Fowler Button 21 5.1.7 Max Inflate/Low Air Loss Button 22 5.1.8 Lock/Alarm Silence 23 5.3 Placing the Patient on the Mattress 24 5.4 Transferring Patient From/To a Service Life 27 6.2 Cleaning the System 27 6.3 Disposal 29 6.4 Storing the System 29 7 Troubleshooting 30 7.1 Troubleshooting
  • Invacare MA65 | Owners Manual - Page 4
    microAIR® MA65 Series 1 General 1.1 Symbols Signal words are used in this manual and apply to hazards or unsafe practices which injury or both. Gives useful tips, recommendations and information of efficient, trouble-free use. THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT
  • Invacare MA65 | Owners Manual - Page 5
    this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as user manuals, service manuals or instruction sheets supplied with this product or optional equipment. DANGER! Risk of Death, Injury
  • Invacare MA65 | Owners Manual - Page 6
    microAIR® MA65 Series DANGER! Risk Of Death, Injury or modification: - DO NOT use this product in any way other than described in this manual. - DO NOT modify the product or any components of the product. WARNING! Risk Of consult the patient's physician before using the mattress system. 6 1148137-H
  • Invacare MA65 | Owners Manual - Page 7
    power cords and outlets for damage. Stop using the product immediately if the cord or outlet is damaged. Contact qualified technician or Invacare for service. WARNING! Risk Of Injury Or Damage To avoid electrical damage, product damage, shock and/or personal injury: - Keep all bed components and
  • Invacare MA65 | Owners Manual - Page 8
    or density could increase risk of entrapment. - The Invacare mattress MUST fit firmly against the bed frame AND bed side rails to prevent patient entrapment. Follow the manufacturer's product instructions. - After any adjustments, repair or service and before use, make sure all attaching hardware is
  • Invacare MA65 | Owners Manual - Page 9
    risk of death or injury from entrapment or falling: - Invacare suggests rails be in the raised or guard position whenever a with oxygen administering equipment. - Do not allow oxygen tent to extend below mattress support level. Installation WARNING! Risk Of Injury Or Damage To avoid serious injury
  • Invacare MA65 | Owners Manual - Page 10
    of the bed. Repair and Service DANGER! Risk of Death, Injury, or Damage To avoid death, injury or damage due to improper maintenance or inspection. Always maintain and inspect equipment per the instructions in this manual. Contact a qualified technician or Invacare if any of the following issues
  • Invacare MA65 | Owners Manual - Page 11
    3 Overview 3.1 Label Location Labels are subject to change without notice. This label is on the back of the unit and also contains the serial number. Overview WARNING! Risk of Injury, Damage or Death Missing or damaged labels may contribute to injury, damage or death. - Ensure labels are present
  • Invacare MA65 | Owners Manual - Page 12
    microAIR® MA65 Series 3.2 Technical Data Electrical Specifications INPUT VOLTAGE AC: INPUT FREQUENCY: CURRENT: MAXIMUM POWER CONSUMPTION: CIRCUIT PROTECTION: MODE OF OPERATION: 110 V-220Vac 50-60 Hz 1A
  • Invacare MA65 | Owners Manual - Page 13
    8 ± 4 mmHg 32 ± 4 mmHg 5, 10, 15, 20 minutes Control unit and mattress have Latex-Free components Overview Mechanical Specifications CONTROL UNIT DIMENSIONS (L X W X H): WEIGHT: POWER CORD: CONNECTION: PACKAGING: AIR FILTER: MA65 15" x 6.5" x 11" 15 lbs 10 - 14 Feet Long, Hospital Grade Three
  • Invacare MA65 | Owners Manual - Page 14
    microAIR® MA65 Series 4 Setup 4.1 Unpacking 1. Remove any loose packing from the carton. 2. Carefully remove all components from the carton. 3. Inspect all components for damage. Contact dealer or Invacare if components are damaged. DO NOT use product. DANGER! Risk of Injury or Damage Continued use
  • Invacare MA65 | Owners Manual - Page 15
    it on the bed frame. Ensure that the hose E is towards the foot end B of the bed. 4. Use the buckle straps F to secure the powered mattress to the bed deck in the following locations: • Head End-Head End Bed Deck A • Foot End-Foot End Bed Deck B • Center-Center of the Bed
  • Invacare MA65 | Owners Manual - Page 16
    microAIR® MA65 Series 4.3 Installing the Bed Side Rails Refer to the Bed Rail Entrapment Risk Notification Guide at www.invacare.com for additional safety information. Refer to the instructions that the hose connecting the control unit to the mattress is routed such that it cannot be stepped on, kinked
  • Invacare MA65 | Owners Manual - Page 17
    pinched, or severed cords: - Ensure that all cord(s) are routed and secured properly. - Keep all moving parts, including the main frame, mattress deck, and all drive shafts, free of obstruction (i.e. blankets/sheets, tubing, wiring, cords, etc.) during operation. 1. Examine the hospital grade power
  • Invacare MA65 | Owners Manual - Page 18
    microAIR® MA65 Series 4.7 to be installed and used according to the EMC information in this manual. This equipment has been tested and found to comply with EMC powered device to a different electrical power circuit. Contact Invacare for any additional EMC information or EMC tables for the device
  • Invacare MA65 | Owners Manual - Page 19
    Usage 5 Usage 5.1 Using the Front Panel DETAIL "B" - MA65 SERIES FRONT PANEL 5.1.1 Power Button 1. To turn the control unit on or off, press and release the Power button A. Once the unit is plugged in,
  • Invacare MA65 | Owners Manual - Page 20
    microAIR® MA65 Series 5.1.2 Alternating Pressure Button 1. Press the Alternating Pressure button B to select between one of four alternating pressure times - 5, 10, 15 and 20 minutes. The alternating pressure times indicate the frequency of deflation or inflation of half of the air the mattress will
  • Invacare MA65 | Owners Manual - Page 21
    from bottoming out. Automatic Wireless Mode When the bed articulates to 45°, the transmitter in the head of the mattress signals the control unit to add 25% more air into the mattress to prevent the patient from bottoming out. Fowler Transmitter Setup 1. Make sure the bed is in the flat position
  • Invacare MA65 | Owners Manual - Page 22
    microAIR® MA65 mattress). Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED. It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient cleaning. • Low Air
  • Invacare MA65 | Owners Manual - Page 23
    Once the low pressure problem is fixed, the control unit resumes operation in the previously set mode. Periodically test alarm by disconnecting the mattress hose. It may take up to 15 minutes for unit to lose enough air for alarm to sound. See troubleshooting section of this manual if alarm fails to
  • Invacare MA65 | Owners Manual - Page 24
    microAIR® MA65 Series 5.3 Placing the Patient on the Mattress 1. Press the Max Inflate button to turn on the control unit to maximum flow. In this mode, the Max Inflate Led lights up. 2. After the mattress is fully inflated, place the patient on the mattress. 3. Ensure that the patient's feet are
  • Invacare MA65 | Owners Manual - Page 25
    pressure. 4. Raise or lower the bed to match the wheelchair height. Refer to the user manual provided with the bed. 5. When the mattress has reached maximum firmness, perform one of the following: • Bed to Wheelchair Transfer - Slide the patient onto the wheelchair. • Wheelchair to Bed Transfer
  • Invacare MA65 | Owners Manual - Page 26
    microAIR® MA65 Series 5.6 Preparing for CPR Procedure 1. Press and hold the tabs A on hose connector B while pulling the hose C from the control unit D. 2. Twist the CPR connector E at the foot end of the overlay F and all air will be evacuated in the torso and chest region of the overlay within 18
  • Invacare MA65 | Owners Manual - Page 27
    6.1 Service Life WARNING! Risk of Injury or Damage Use of the product beyond this time period may cause product damage and injury. - This product has an expected lifetime of one (1) years when used in accordance with safety instructions, maintenance intervals and correct use stated in this manual
  • Invacare MA65 | Owners Manual - Page 28
    microAIR® MA65 Series avoid damage, the following steps are recommended for cleaning the mattress cover: a. To disinfect the cover: i. Using a clean below 120°F). iii. Add detergent and disinfectant according to the manufacturer's instructions. iv. Remove excess water. v. Set the dryer to the lowest
  • Invacare MA65 | Owners Manual - Page 29
    outlet. 2. Clean the system. Refer to 6.2 Cleaning the System, page 27. 3. Disconnect the air hose connector(s) from the control unit and allow air to vent from the mattress. 4. Gently roll up the mattress with minimal handling and agitation. 5. Ensure the cover surface is inside the roll. 6. Store
  • Invacare MA65 | Owners Manual - Page 30
    microAIR® MA65 Series 7 Troubleshooting 7.1 Troubleshooting PROBLEM Mattress not inflating Not alternating properly No power 30 CAUSE SOLUTION Mattress hose disconnected Connect hose connectors and lock them in place Air hose kinked or split Unkink hose or replace split hose Major leak in the
  • Invacare MA65 | Owners Manual - Page 31
    power unit Power cord disconnected No power in the power source Power outage Blown fuse Mattress hose disconnected Troubleshooting SOLUTION Replace battery Replace fowler Refer to Folwer transmitter setup in this manual Connect power cord to the power source Check power source has power and turn it
  • Invacare MA65 | Owners Manual - Page 32
    microAIR® MA65 Series 8 also have other legal rights which vary from state to state. Invacare warrants the mattress and cover when purchased new and unused to be free from defects SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS. IN ADDITION, THE
  • Invacare MA65 | Owners Manual - Page 33
    CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL
  • Invacare MA65 | Owners Manual - Page 34
    Notes
  • Invacare MA65 | Owners Manual - Page 35
    Notes
  • Invacare MA65 | Owners Manual - Page 36
    Invacare Corporation USA One Invacare Way Elyria, Ohio USA 44035 800-333-6900 1148137-H 2017-05-02 *1148137H* www.invacare.com Canada 570 Matheson Blvd E Unit 8 Mississauga Ontario L4Z 4G4 Canada 800-668-5324 Making Life's Experiences Possible®
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

microAIR® MA65 Series
MA65, MA65B42, MA65B48, MA65RSR
en
Alternating Pressure Low Air Loss Mattress
User Manual
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.