Fluke T6-600 Product Manual

Fluke T6-600 Manual

Fluke T6-600 manual content summary:

  • Fluke T6-600 | Product Manual - Page 1
    T6-1000/T6-600 Electrical Tester Service Information February 2018 ©2018 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. Specifications are subject to change without notice. MyFlukeStore Shop for Fluke products online at: www. .com 1.888.
  • Fluke T6-600 | Product Manual - Page 2
    . The warranty period is two years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any
  • Fluke T6-600 | Product Manual - Page 3
    7 FieldSense Voltage Function 9 Resistance and Continuity Functions 9 Low Battery Indicator 10 Battery Door Circuit 10 Maintenance 11 How to Clean 11 Battery Replacement 11 Replacement Parts 12 i MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke T6-600 | Product Manual - Page 4
    you are qualified to do so. The information provided in this manual is for the use of qualified personnel only. Safety Information A Warning identifies hazardous conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the
  • Fluke T6-600 | Product Manual - Page 5
    T6-1000/T6-600 Service the Product behind the tactile barrier. See the Quick Reference Guide for this Product. • To use the FieldSense Ground Contact, the Product. • Use only specified replacement parts. For safe operation and maintenance of the Fluke products online at: www. .com 1.888.610.7664
  • Fluke T6-600 | Product Manual - Page 6
    VOLTAGE. Risk of electric shock. Consult user documentation. FieldSense Measurement: Fluke voltage/current sensing technique. T Double is applicable to test and measuring circuits connected to the distribution part of the building's low-voltage MAINS installation.  Measurement Category
  • Fluke T6-600 | Product Manual - Page 7
    T6-1000/T6-600 Service T6-600 CAT III 600 V T6-1000 CAT IV 600 T6-1000 Range Model T6-600 Range Resolution Accuracy[1] FieldSense ac voltage true-rms No 1000 V 600 true-rms Yes 1000 V 600 V 1 V ±(1.5 % + 2 counts) 45 Hz to 66 Hz Volts dc Yes 1000 V 600 V 1 V ±(1 % for user wearing
  • Fluke T6-600 | Product Manual - Page 8
    a 1-year cycle for the Performance Tests. If the Product fails any test, return the Product to a Fluke Service Center or Authorized Service Partner. Before you start the Performance Tests, make sure that all test equipment has adequate warm-up time and is allowed to stabilize in an
  • Fluke T6-600 | Product Manual - Page 9
    T6-1000/T6-600 Service Information Voltage Function Perform these tests as follows: 1. Set the calibrator NORMAL output to the voltage listed in Table 5, Step 1. Table 5. AC Voltage Tests T6-600 T6 steps in Table 6. MyFlukeStore 6 Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke T6-600 | Product Manual - Page 10
    the SHIFT (yellow) button on the Product to switch to AC Current mode and Frequency mode (T6-1000 only). 2. Connect the 10/50-turn coil to the calibrator AUX HI and AUX LO jacks fork on the 10/50-turn coil. See Figure 1. MyFlukeStore Shop for Fluke products online at: www. 7 .com 1.888.610.7664
  • Fluke T6-600 | Product Manual - Page 11
    T6-1000/T6-600 Service Information 6. Verify that the Product reading is within the display limits. 7. Apply the test values for A 96.7 A to 103.3 A 96.7 A to 103.3 A 174.3 A to 185.7 A 174.3 A to 185.7 A 174.3 A to 185.7 A MyFlukeStore 8 Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke T6-600 | Product Manual - Page 12
    Proving Unit (Fluke P/N 4910310) and a Fluke 8846A Multimeter. 1. Set the 8846A to the AC Voltage measurement mode (ACV button) and range manually to the 5 6 7 Calibrator Output 26 Ω 300 Ω 400 Ω 800 Ω 1600 Ω ----- T6-600 Display Limits/ Beeper Beeper On Beeper Off 394 Ω to 406 Ω 790 Ω to 810 Ω
  • Fluke T6-600 | Product Manual - Page 13
    T6-1000/T6-600 Service Information Low Battery Indicator To verify the correct operation of the low battery indicator: 1. Remove the Tester the battery door from the test instruments and reinstall it onto the Tester. MyFlukeStore 10 Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke T6-600 | Product Manual - Page 14
    2. Battery Replacement 2. Remove the battery door. 3. Replace the batteries. 4. Attach the battery door and battery door screw. Do not overtighten the screw. MyFlukeStore Shop for Fluke products online at: www. 11 .com 1.888.610.7664
  • Fluke T6-600 | Product Manual - Page 15
    T6-1000/T6-600 Service Information Replacement Parts Replacement parts and accessories are listed in Table Table 10. Replacement Parts 4 3 1 2 Fluke Part Ref. Description or Model Number  Battery Door 4944370  Battery Door Screw 1618579  Test Lead Assembly (T5-RLS). Replace only with
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15

February 2018
©2018 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
Specifications are subject to change without notice.
T6-1000/T6-600
Electrical Tester
Service Information
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com